Container Closure Integrity (CCIT) vs Sterility Testing

By Lim Xu, Export Manager

In CCIT vs sterility testing, the two tests answer different questions. A sterility test checks whether viable microorganisms are present in the samples tested, while container closure integrity testing (CCIT) checks whether the package can keep contamination out. Sterile products need both, because neither result proves the other.

What does a sterility test tell you?

A sterility test looks for viable microorganisms in samples taken from a batch, and under USP <71> the inoculated media are incubated for at least 14 days. The harmonized pharmacopoeial methods are USP <71>, Ph. Eur. 2.6.1 and JP 4.06, which ICH recommends as interchangeable in its regions under stated conditions.

Drugs and biologics labeled sterile must be tested for sterility batch by batch, as FDA notes in its 2008 guidance on container closure integrity testing. The same guidance sets out the known limits of the test.

  • It only detects viable microorganisms present at the time of the test.
  • It only detects organisms that can grow in the specified culture media.
  • Contamination introduced during testing can give a false positive.
  • It always destroys the samples, so the same units cannot be examined again.

What does container closure integrity testing tell you?

Container closure integrity testing checks whether the container and closure system can keep its barrier against microbes. FDA treats sterility as a stability characteristic, so a sterile product is expected to stay free of viable contamination for its whole shelf life.

FDA lists properly validated physical, chemical and microbiological tests that can do this job. Examples include bubble tests, pressure or vacuum decay, trace gas leak tests, dye penetration and microbial challenge.

USP <1207> groups leak test technologies as deterministic or probabilistic. Microbial challenge and dye ingress sit in the probabilistic group, because their results carry some uncertainty. That makes them hard to validate quantitatively for detecting critical leaks.

Questions about a model or a test? Email Amy Zhao, Export Manager, at [email protected].

Why can't one test replace the other?

CCIT cannot stand in for the release sterility test. FDA's guidance says container and closure integrity tests cannot demonstrate a product's initial sterility.

The reverse also holds. Because of the limits listed above, FDA does not recommend sterility tests for confirming continued sterility through shelf life. An integrity test may detect a breach in the package before the product is contaminated.

Where does each test fit in a sterile product program?

The table below sets the two tests side by side on the points that decide where each one is used.

PointSterility testContainer closure integrity test
Question answeredAre viable microorganisms present in the samples tested?Can the package keep contamination out?
Batch releaseRequired for products labeled sterileCannot show initial sterility, so it does not replace the release test
Stability programNot recommended for confirming continued sterilityCan replace the sterility test at later time points, such as annually and at expiry
SamplesAlways destroyedSome methods leave samples for other stability tests
Time to resultNeeds days of incubation in culture mediaOften shorter, depending on the method

In practice the two run side by side. Sterility testing covers each batch at release, and a validated integrity test covers the package through its shelf life.

How is a CCI test chosen for a package?

A CCI test only counts once it has been validated for that container and closure system. FDA treats a method as adequately validated when studies prove it can detect a breach in integrity.

USP <1207> separates a leak test technology from a leak test method. After a technology is chosen, method development has to show that it detects a critical leak in the specific product package, using defined test parameters. No single test suits every package.

For flexible packages, our MFY-01B sterile package seal tester runs the ASTM D3078 bubble emission leak test. This method finds gross leaks in flexible packages that contain headspace gas, and small leaks may not be detected.

The MFY-01 semi-auto vacuum seal tester runs the same ASTM D3078 method.

Frequently asked questions

Can preservative effectiveness testing replace a CCI test?

No. FDA does not accept preservative effectiveness tests as an alternative for monitoring container and closure integrity. They only show how the preservative performs against a panel of five test organisms, not whether contamination is present.

Does a gas-tight package also keep microbes out?

Generally, yes. A container that is gas-tight will usually also be tight against microbial ingress. That is why products packed under inert gas or partial vacuum set a stricter integrity requirement than sterility alone.

Can a stability program keep its sterility tests?

Yes. FDA's guidance says manufacturers that already run sterility testing in a stability protocol may continue to do so. The integrity test route is offered as an alternative, not as an added requirement.