Bursting Strength Tester Calibration and What Foils Can Check
By Lim Xu, Export Manager
Bursting strength tester calibration under ISO 2758 and ISO 2759 covers the pumping rate, the maximum pressure reading and any clamping pressure indicator. Aluminium foils of known burst value help confirm that the instrument works, but both standards advise against using them as calibration standards.
What calibration covers on a burst tester
Both ISO methods expect the tester to have test points, or room to fit them, for three checks. These are the fluid pumping rate, the maximum pressure measurement and display, and the clamping pressure indicator where one is fitted.
Calibration is required before the tester is first used. After that, the standards ask for intervals short enough to hold the specified accuracy, rather than a fixed period.
Incorrect calibration of the pressure measuring system heads the list of main error sources in both standards. A drifting gauge or transducer shifts every result the tester produces.
Why foils are a function check, not a calibration
ISO 2758 and ISO 2759 mention aluminium foils of various thicknesses that may be available as test pieces of known burst value. Both standards describe them as a useful means of checking the overall function of an instrument.
The standards give the reason: foil behaves differently from paper under stress, so foils should not be used as calibration standards.
In practice, a foil result can show that something has changed, but not how far the pressure reading is out. Run foil checks between calibrations, record each result, and treat a shift as a reason to calibrate the pressure system.
Questions about a model or a test? Email Amy Zhao, Export Manager, at [email protected].
Calibrating the pressure measuring system
The pressure system may work on any principle. Measurement and display together must hold ±10 kPa or ±3% of the value, whichever is larger.
The ±10 kPa limit governs up to about 333 kPa, and the 3% limit governs above that. At 1,000 kPa, for example, the allowance is ±30 kPa.
Speed of response matters as well. The displayed maximum must sit within ±3% of the true peak, measured with the kind of calibration system set out in Annex D.
Pumping rate, clamping and diaphragm checks
The paper and board methods set different values, so check the tester against the method it is used for. The table below compares the main check values in the two ISO methods.
| Check | ISO 2758 (paper) | ISO 2759 (board) |
|---|---|---|
| Pumping rate | 95 ml/min ± 5 ml/min | 170 ml/min ± 15 ml/min |
| Diaphragm bulge | 30 kPa ± 5 kPa at 9.00 mm | 170 to 220 kPa at 10 mm; 250 to 350 kPa at 18 mm |
| Clamping pressure | Known pressure up to 1,200 kPa | 700 to 1,200 kPa |
| Pressure accuracy | ±10 kPa or ±3% | ±10 kPa or ±3% |
In both methods, the clamp openings must be concentric to within 0.25 mm under load, and Annex B gives the way to check the clamps. Bulge heights carry a tolerance of ±0.2 mm. A clamping indicator should preferably show the actual clamping pressure, which can be worked out from clamping force and clamp area.
Our step by step burst test guide covers the checks to make before a test session.
Calibration records
A short record for each calibration makes drift visible over time. A practical record includes these items:
- Date, instrument and the method it serves, such as ISO 2758, ISO 2759, TAPPI T 810 or TAPPI T 403
- Pumping rate found
- Pressure readings against the reference, before and after any adjustment
- Clamping pressure indicator check, where one is fitted
- Reference instrument used and its own calibration status
- Foil check results and diaphragm changes since the last calibration
In North America, TAPPI T 403 is the paper method and TAPPI T 810 the method for single and double wall corrugated board. Their public summaries do not describe calibration, so take the details from the full method your customer names.
Frequently asked questions
Does an overloaded pressure sensor need recalibration?
Yes. ISO 2758 and ISO 2759 both say that a pressure sensor accidentally taken beyond its rated capacity shall be recalibrated before further use. Log the event and do not run further tests until the sensor is recalibrated.
Can the pressure sensor be calibrated away from the tester?
It is better not to. Both standards prefer the sensor calibrated where it normally sits, ideally while still fitted to the tester. Bench calibration is a fallback when that is not possible.
Talk to our factory team
We sell direct from our factory, so you work directly with our factory team. Questions about a model or a test? Email Amy Zhao, Export Manager, at [email protected].
Related articles
- Box Compression Test Formula and the McKee EquationThe box compression test formula most labs use is the simplified McKee equation. See each term, a worked ECT to BCT example and when to test boxes instead.
- Pouch Burst Test for Flexible and Medical PackagingA pouch burst test inflates a sealed package until it fails. Compare ASTM F1140 and F2054, burst and creep tests, and how to read where the pouch opened.
- Bursting Strength Tester HS Code by CountryHow a bursting strength tester moves from HS subheading 9024.80 to national tariff lines, how binding rulings work, and what paperwork customs needs.
- How to Use a Bursting Strength Tester Step by StepHow to use a bursting strength tester on board: condition the samples, check the diaphragm and clamp, run 20 valid bursts, then read and report the result.