Pressure Decay Leak Test for Packages and Rigid Containers

By Lim Xu, Export Manager

A pressure decay leak test fills a package or rigid container with air or another gas, lets the pressure settle, then watches for a drop during a hold. A drop that is larger than the limit you set is treated as a leak.

How a pressure decay leak test works

The instrument raises the internal pressure to a set value and isolates the system from the supply. After a short stabilize step, it records pressure through the hold. If gas escapes through a seal or a wall, the pressure falls.

ASTM F2095 covers this method for nonporous film, foil or laminate pouches and for foil-sealed trays, empty or holding a solid product. The official abstract lists two setups: Test Method A, with the package free to inflate, and Test Method B, with rigid restraining plates against the walls.

F2095 also notes that these methods require a leak-tight entry into the package to supply the gas, usually air or nitrogen. That entry is why the standard treats the methods as destructive to the pack. Closed rigid containers may use a similar fill, stabilize and hold cycle on a fixture made for that container. Which setup fits a given pack is something to confirm with us first.

Pressure decay vs vacuum decay

The two methods look at pressure change from opposite sides. A pressure decay leak test raises pressure inside the pack and watches for a fall. A vacuum decay leak test draws a vacuum around or on the pack and watches for a rise, as ASTM F2338 describes.

Choose pressure decay when you can pressurize the pack safely and want a quantitative hold result. Vacuum decay is often used when you should not inflate the pack, or when the method of record is a vacuum method.

Flexible packs change volume as they inflate. F2095 Method A requires the package to reach a stable volume, in other words to stop stretching, before the hold is read. Method B uses plates to limit that volume so the reading is less affected by stretch.

The test cycle of fill, stabilize, hold and vent

A typical cycle has four steps: fill to the set pressure, stabilize, hold and measure, then vent. The fill must be slow enough not to shock the seal. The vent should be controlled so the pack is not damaged on the way down.

Why stabilization time matters

During fill, the pack, the gas and the fittings are still settling. A flexible wall may still be stretching. Gas in the lines may still be cooling after compression. If you start the hold too soon, that settling looks like a leak.

The stabilize step is there so the hold starts from a steady baseline. If results jump from test to test, lengthen the stabilize time before you tighten the pressure limit.

Temperature and volume effects

Pressure in a closed volume changes with temperature. A warmer room, a pack that has just come off a line, or a draft across the fixture can move the reading even when the seal is sound.

Volume changes have the same effect. If a flexible pack creeps during the hold, the pressure falls without a leak path. Restraining plates, a longer stabilize step and a stable room all reduce that background change.

F2095 points out that the instrument reading is in pressure units. Any conversion to a leak rate is a calculation on top of that reading, and the gauge itself has to be calibrated in pressure units. This article does not give a conversion formula.

Setting acceptance criteria from a reference leak

Limits do not come from a generic number in a catalogue. They come from packs you already trust, and from a reference leak you can repeat.

F2095 notes that calibrated hole sizes or orifices may be used to judge how small a leak the setup can see. Run known-good packs to learn the background decay of a sound pack at your volume and pressure. Then run packs with a reference leak and set the hold limit between those two groups.

Record the pack type, the fill pressure, the stabilize and hold times, whether plates were used, and the room temperature. A limit that works on a small tray will not transfer automatically to a larger pouch, because the same leak changes the pressure less in a larger volume.

Labs that already run a bubble leak test can keep it as a visual screen and add pressure decay when they need a measured hold. The two results are complementary, not substitutes.

If you want to see a pressure decay leak tester for this work, tell us the pack type and the method you need to follow. Whether a given setup matches a named standard is something to confirm with us first.

Questions about a model or a test? Email Amy Zhao, Export Manager, at [email protected].

Where pressure decay fits next to bubble and burst tests

A pressure decay leak test gives a quantitative hold result and can be set against a reference leak. It is not a visual bubble check, and it is not a burst test.

A bubble leak test shows gross leaks as bubbles under liquid, usually by ASTM D3078 or ASTM F2096. It is simple to interpret, but small leaks can be missed. A burst or seal-strength test raises pressure until the pack fails. That is a different question from whether a pack holds a set pressure through a hold.

Use bubble testing for a fast visual screen, pressure decay when you need a measured hold, and burst testing when you need the pressure at failure. Many labs run more than one of these, because they do not replace each other.

Frequently asked questions

Is a pressure decay leak test the same as a burst test?

No. Pressure decay holds a set pressure and looks for a drop. A burst test raises pressure until the pack fails and records that failure pressure.

Does ASTM F2095 apply to rigid bottles?

The published scope of F2095 is flexible pouches and foil-sealed trays. A rigid bottle can still be tested with a fill, stabilize and hold cycle, but that setup is not the same as the F2095 pouch methods. Confirm the method with us before you write it into a specification.

What should I send with a request for a quotation?

Tell us the pack or container type, whether you need a restrained or free setup, and the method you have to follow. We can then say which configuration to consider and send you a quotation.

Talk to our factory team

We sell direct from our factory, so you work directly with our factory team. Questions about a model or a test? Email Amy Zhao, Export Manager, at [email protected].

Related articles

Let's Talk

We're here to help you with your needs. Let's talk about how we can help you.

View our privacy policy.

© 2026 KHT Instruments | All rights reserved.